Estimation of Mean Nd:YAG Laser Capsulotomy Energy Levels for Different Grades of Posterior Capsular Opacification: A Comparative Study.

Surendra Wadgaonkar, Professor & Head, Department of Ophthalmology, A.C.P.M. Medical College, Dhule, Maharashtra 424002, India , Sonam S. Rathod1 , Surendra P. Wadgaonkar2 , Sanjay V. Vaghmare3 , Preeti A. Patil4

Author Information

Licence:




Ophthalmology and Allied Sciences 4(2):p 131-138, May - August 2018. | DOI: DOI: http://dx.doi.org/10.21088/oas.2454.7816.4218.15

How Cite This Article:


Timeline

Received : N/A         Accepted : N/A          Published : N/A

Abstract

Background: Patients who have PCO with significantly reduced visual acuity need posterior capsulotomy to improve vision. Nd:YAG laser for posterior capsulotomy is a widely used, safe, noninvasive, and effective procedure with minimal complications. The aim of this study is to estimate mean energy levels for Nd:YAG laser capsulotomy in various subtypes of PCO (membranous, fibrous,and fibro-membranous). Setting And Design: This is a prospective comparative study at a rural based, tertiary level hospital. Methods And Materials: This study has been performed at our institute between September 2013 and September 2015 after taking clearance from the ethics committee. 100 eyes of 100 patients with PCO were considered for Nd:YAG laser capsulotomy after minimum period of 6 months following uncomplicated extracapsular cataract extraction.Statistical analysis was done by the unpaired t-test using SPSS software. Results: In our study, the mean initial and mean summated laser energy levels were different for different subtypes of PCO. There was a significant difference in mean initial laser energy levels for different subtypes of PCO (p<0.001). Mean initial laser energy level was minimum for membranous PCO (1.4 mJ) followed by fibro-membranous PCO (2.2 mJ) and maximum for fibrous PCO (3.2 mJ), whereas the mean summated energy level for membranous was 25.22 mJ, for fibrous 51.48 mJ and for fibro-membranous was 70.08 mJ. The mean summated laser energy level was maximum

for fibro-membranous PCO despite having a comparatively lower mean initial energy level (2.2 mJ). It was significantly higher for bag-sulcus fixated IOLs (75.67 mJ) and sulcus fixated IOLs (55.76 mJ) than bag fixated IOLs (52 mJ) (ANOVA=0.035). Conclusion: The mean initial laser energy level and the mean summated laser energy level is different for different subtypes of PCO. The latter is higher for fibromembranous
type of PCO than fibrous and membrane 
 
Keywords: Laser Capsulotomy; Posterior Capsular Opacification (Pco); Mean Energy Levels; Surgical
Capsulotomy.


References

No records found.


About this article


Cite this article


Licence:




Received Accepted Published
N/A N/A N/A

DOI: DOI: http://dx.doi.org/10.21088/oas.2454.7816.4218.15

Keywords


Article Level Metrics

Last Updated

Sunday 26 July 2026, 12:21:43 (IST)


579

Accesses

1
66
00

Citations


NA
NA
NA

Download citation


Article Keywords


Keyword Highlighting

Highlight selected keywords in the article text.


Timeline


Received N/A
Accepted N/A
Published N/A

licence



Access this article



Share