Accuracy of Point of Care Blood Sugar Measurement in Patients Admitted to Intensive Care Unit with Shock and on Vasopressor Support

Suresh Kumar N , Naveed Abrar1 , Suresh Kumar N2 , Dinesh K3 , Sumanth T4 , Waseem Anjum5

Author Information

Licence:




Indian Journal of Anesthesia and Analgesia 6(3):p 737-742, May-June 2019. | DOI: http://dx.doi.org/10.21088/ijaa.2349.8471.6319.6

How Cite This Article:

Naveed Abrar, Suresh Kumar N, Dinesh K et al., Accuracy of Point of Care Blood Sugar Measurement in Patients Admitted to Intensive Care Unit with Shock and on Vasopressor Support. Indian J Anesth Analg. 2019;6(3):737-741.


Timeline

Received : N/A         Accepted : N/A          Published : N/A

Abstract

Background and Aim: Point of care (POC) glucometry is most widely used method for monitoring blood sugar levels in the Intensive care unit (ICU). But, the accuracy of bedside glucometry using capillary blood is controversial in patients with shock. We aimed to compare the accuracy of POC glucometry with laboratory blood glucose measurements in patients with shock and requiring Noradrenaline infusion. Methods: This is Prospective, Observational study done on 100 consecutive patients admitted to ICU with shock categorized into two groups of 50 each, requiring two different doses of Noradrenaline infusion (Group A: 5 mcg/min and Group B: 10 mcg/min). POC capillary, POC venous and laboratory venous blood sugar measurements were simultaneously done in each patient. Results: ANOVA test was performed which showed that there was no statistical significance between the blood samples and also between the two groups (F value 0.022, p value 0.978). In Group A, stronger correlation was found between POC venous and lab venous blood sugar measurements (r=0.886, p<0.01) whereas in Group B, stronger correlation was found between POC capillary and lab venous blood sugar measurements (r=0.897, p<0.01) Conclusion: POC glucometer blood sugar estimation using capillary blood quite accurately correlates with the POC glucometry using venous blood and the laboratory values, even in patients with Noradrenaline infusion, provided all the confounding factors are eliminated before measuring blood sugar levels.


References

No records found.


About this article


Cite this article

Naveed Abrar, Suresh Kumar N, Dinesh K et al., Accuracy of Point of Care Blood Sugar Measurement in Patients Admitted to Intensive Care Unit with Shock and on Vasopressor Support. Indian J Anesth Analg. 2019;6(3):737-741.


Licence:




Received Accepted Published
N/A N/A N/A

DOI: http://dx.doi.org/10.21088/ijaa.2349.8471.6319.6

Keywords

Blood glucose monitoring; Point of care glucometer; Shock.

Article Level Metrics

Last Updated

Tuesday 14 July 2026, 09:06:25 (IST)


5690

Accesses

5
679
00

Citations


NA
NA
NA

Download citation


Article Keywords


Keyword Highlighting

Highlight selected keywords in the article text.


Timeline


Received N/A
Accepted N/A
Published N/A

licence



Access this article



Share